%A Ruiz-Gonzalez, Y. %A Rodriguez-Ledesma, S. %A Paz-Viera, J. E. %A Lorenzo-Ginori, J. V. %D 2020 %T Iso-Dose Map Generation and Dose-Area Product Calculation through Digital Image Processing of Scanned Irradiated Radiochromic Films %B 2020 %9 %! Iso-Dose Map Generation and Dose-Area Product Calculation through Digital Image Processing of Scanned Irradiated Radiochromic Films %K %X Iso-dose maps creation and kerma-area products calculation are important procedures in radiotherapy and medical imaging. In this work, a method is presented to obtain iso-dose maps through digital processing of the images obtained from radiochromic films, which are employed to estimate the dose absorbed by the patient. The kerma-area product in selected regions of interest (ROIs) in the iso-dose map was then evaluated. The images were obtained by scanning the irradiated films using a commercial flatbed scanner. The iso-dose areas were extracted by means of an image segmentation algorithm, while calibration for a particular film to obtain a sensitometric curve relating dose to film darkening, was also made. A software interface allowed introducing the scanned image for the estimation of kerma-area products by defining the ROI interactively. The methods developed in this work allowed to implement a software application to obtain the iso-dose maps and kerma-area product in selected ROIs. %U http://revistacubanadefisica.org/index.php/rcf/article/view?path= %J Revista Cubana de FĂ­sica %0 Journal Article %& 101 %P 7 %V 37 %N 2 %@ 2224-7939 %8 2020-12-13 %7 2020-12-15